Standard
features:
- 3 sizes 108, 153 and 246 litres
- temperature range +20°C up to +45°C
- CO2 adjustable from 0% up to 20% with a
setting accuracy of 0,1%
- 2 connections with automatic change-over device for
gas bottles
- CO2 infra-red absorption system with
automatic zeroing
- large-area heating on all sides
- integrated sterilisation stage at +160°C (set point
wait dependent) for 2 hours -sterilisation (NOT DISINFECTION)
without removal of sensors
- doubled overheat safety by microprocessor overheat
controller, adjustable (TWW) and temperature limiter TB safety device
class 3.1 and class 1 according to DIN 12 880
- stainless steel 1.4301, interior
electropolished
- serial interface RS232 incl. MEMMERT-software
„Celsius 2000“ or optional RS485 (no extra charge)
- parallel printer interface for connection of an
EPSON -compatible printer for documentation in accordance with GLP (Good
Laboratory Practice)
- works calibration certificate for one temperature:
+37°C
|